Description :
INFINITY-CTW is an accurate conductivity and temperature meter making use of 7-electrode sensor with a wiper to avoid bio-fouling. This 7-electrode generates an electric current only inside of the conductivity cell, minimizing external influences improving data quality. Thus, precise salinity measurements are possible under bio-foul forming conditions. The compact design is suitable to be deployed/integrated with various sites/platforms
Calibration Information :
Calibration Frequency :
Biannually
Calibration Standard Method :
Calibration equation is determined from fifth order regression of samples of the reference temperature against A/D values. Samples are taken at approximately 0, 5, 10, 15, 20, 25, 30, and 35 °C.
Calibration Certificate :
As of Certificate :
04/11/2022